Menu Close

MS-60 Albedometer

271,075.62

Make:  Eko    Model:  MS-60
✅ Albedometer effectively measures albedo and total reflectance across various applications, including meteorology and solar power plants.
✅ Measurand : albedo and total reflectance.
✅ Sensor Type : Thermophile (ISO 9060:2018).
✅ Signal Output : Analog Output, , MODBUS 485RTU, SDI-12, 4-20mA and 0-1V (external 100 Ω precision shunt resistor).
✅ Supplied With : 10m (Optional lengths 20m, 30m, 50m)
                
Compare
Category:

The MS-60 Albedometer, with double-dome construction for low offsets and cosine errors,  effectively measures albedo and total reflectance across various applications, including meteorology and solar power plants with bifacial PV modules.

The MS-60 Albedometer consists of two MS-60 ISO 9060:2018 Class B Pranometers with a double dome construction for lower offsets and cosine errors. Combined with the MS-Albedo Kit, plus glare screen, the MS-60 Albedometer effectively measures albedo and total reflectance across various applications, including meteorology and solar power plants with bifacial PV modules.

MS-60 Pyranometers have a 5-year warranty and are manufactured with strict quality controls and performance evaluation. EKO provides calibration compliant with the international standards defined by ISO/IEC17025/9847 and a 2-year recommended recalibration interval.

Make

Eko (Japan)

Model

MS-60 Albedometer

ISO 9060:2018

Class B

ISO 9060:1990

(First Class)

Sub-category "Spectrally flat"

Compliant

Sub-category "Fast response"

Not Compliant

Output

Analog (mV)

Response time 95%

< 18 Sec.

Zero offset A - Thermal Radiation (200W/m²)

± 5 W/m²

Zero offset B - Temperature change (5K/hr)

± 2 W/m²

Zero offset C - Total zero off-set

± 7 W/m²

Non-stability (change/year)

± 1.5 %

Non-linearity at 1000W/m²

± 1 %

Directional Response (at 1000W/m² | 0 to 80°)

± 18 W/m²

Spectral Error

± 0.2 % (Spectral selectivity ± 3%)

Temperature response -20°C to 50°C

± 4 %

Tilt Response (0-90° | 1000W/m²)

± 1 %

Wavelength Range (nm)

285 to 3000 (50% points)

Irradiance range (W/m²)

0 to 2000

Signal Output

Analog

Operating temperature

-40 to +80 °C

Ingress Protection

IP 67

Calibration traceability / uncertainty

ISO 17025 / WRR / < 0.7% (k = 1.96)ISO 17025 / WRR / < 0.7% (k = 1.96)

Standard Cable Length

10m (Optional lengths 20m, 30m, 50m)